Human IL-8/Interleukin-8/CXCL8 ELISA Kit PicoKine®

Cat# EK0413

Size : 96wells/kit,withremovablestrips.

Brand : Boster Bio

Contact local distributor :


Phone : +1 850 650 7790

Product Info Summary

SKU: EK0413
Size: 96 wells/kit, with removable strips.
Reactive Species: Human
Application: ELISA
Sample Types: cell culture supernatants, serum and plasma (heparin, EDTA, citrate).

Product Name

Human IL-8/Interleukin-8/CXCL8 ELISA Kit PicoKine®

SKU/Catalog Number

EK0413

Size

96 wells/kit, with removable strips.

*Question: How many samples can I assay/run in this kit?

Description

Human IL-8/Interleukin-8/CXCL8 ELISA Kit PicoKine® (96 Tests). Quantitate Human CXCL8 in cell culture supernatants, serum and plasma (heparin, EDTA, citrate). Sensitivity: 1pg/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.

Storage & Handling

Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)

Cite This Product

Human IL-8/Interleukin-8/CXCL8 ELISA Kit PicoKine® (Boster Biological Technology, Pleasanton CA, USA, Catalog # EK0413)

Clonality of Antibodies

See Datasheet for details

Standard Protein

Expression system for standard: E.coli; Immunogen sequence: S28-S99

Sensitivity

<1 pg/ml

Assay Range

15.6 pg/ml - 1,000 pg/ml

Standard Dilution Instructions

serial dilution instructions image

See datasheet of EK0413 for more details

Cross-reactivity

There is no detectable cross-reactivity with other relevant proteins.

Reactive Species

EK0413 is reactive to CXCL8 in Human samples

Validated Sample Types

cell culture supernatants, serum and plasma (heparin, EDTA, citrate).

Application Guarantee

EK0413 is guaranteed for ELISA in Human by Boster Guarantee

See how Boster Bio validate our ELISA kits: ELISA Validation Information

Background of CXCL8

Interleukin-8, also called neutrophil-activating peptide-1 or SCYB8, is a tissue-derived peptide secreted by several types of cells in response to inflammatory stimuli. Monocyte-derived neutrophil chemotactic factor (MDNCF/IL-8, suggested gene symbol IL8) is a cytokine that chemoattracts and activates neutrophils. IL-8 is produced and released from human adipose tissue and from isolated adipocytes in vitro, which may indicate that IL-8 from adipose tissue could be involved in some of the obesity-related complications. The MDNCF/IL-8 gene is placed on the human gene map at position 4q12-q21. This is the same location where at least three other members (platelet factor 4, melanoma growth stimulatory activity, and interferon-gamma induced factor) of the platelet factor 4 gene superfamily reside. Human IL-8 consists of 99 amino acids in precursor form and 79 amino acids in mature form.

Kit Components

Catalog Number Description Quantity
EK0413-CAP Anti-Human CXCL8 Pre-coated 96-well strip microplate 1
EK0413-ST Human CXCL8 Standard 2 vials, 10 ng/tube
EK0413-DA Human CXCL8 Biotinylated antibody (100x) 100ul
AR1103 Avidin-Biotin-Peroxidase Complex (100x) 100ul
AR1106-1 Sample Diluent 30ml
AR1106-2 Antibody Diluent 12ml
AR1106-3 Avidin-Biotin-Peroxidase Diluent 12ml
AR1104 Color Developing Reagent (TMB) 10ml
AR1105 Stop Solution 10ml
AR1106-5 Wash Buffer (25x) 20ml
PLA-SEA Adhesive plate sealers 4

*The kit components are not available for individual purchase.

Materials Required But Not Included In Kit

  • Microplate Reader capable of reading absorbance at 450nm.
  • Incubator.
  • Automated plate washer (optional).
  • Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions.
  • Multichannel pipettes are recommended for large amount of samples.
  • Deionized or distilled water.
  • 500ml graduated cylinders.
  • Test tubes for dilution.

Validation Standard Curve O.D. At 450nm

Concentration (pg/ml)015.631.262.51252505001000
O.D.0.0220.0660.1060.1990.3620.6811.252.095

Data Example Images

Intra Assay Consistency & Inter Assay Consistency

We measured random samples of Human IL-8/Interleukin-8/CXCL8 ELISA Kit PicoKine® within the same batch/lot to ensure the consistency of the kits' performances. ELISA intra assay consistency is measured using wells from the same plate/assay kit. ELISA inter assay consistency is measured using wells from different plates from the same batch production/lot.

Intra-Assay PrecisionInter-Assay Precision
Sample123123
n161616242424
Mean (pg/ml)16662341663230
Standard deviation0.722.8314.270.83.8417.48
CV (%)4.5%4.3%6.1%5%6.1%7.6%

You might also be interested by the following products:



Cat#
Description
Cond.
Price Bef. VAT